1. Dirty DRV(s).
2. Clean or replace DRV(s).
3. Make DRV Diagnosis test to assure correctness of operation
Requires to be ordered and Machine to be Activated.
1. Replace Motor and other failed parts.
2. Upgrade to Manual Hood Operation. Retrofit upgrade kit available. Recommended.
The button should be swiched ON when the testing tank is empty or the flush/wast tank are full
Check Level Sensor
1. Donwload the Software copy to a USB and connect it to the unit.
2. Press USB update Video Instructions Available
Update via usb
1. PCB Board output K16 failure.
2. New upgrade available
1. Assure the CFL step was not Skipped.(Indicated on the test report too)
2. Assure injector Warm-up was Not Skipped. (Indicated on the test report too)
3. Check In-Line filters and replace periodically
4. Replace the Calibration Oil with fresh/new one.(Check the timers)
5. Assure Not to use the flush adpater for measurement. Only Use of RSP or iPSC (Spray Chamber).
6. Assure the Pressure is correct by doing a verification (MVC1) using a Master hight pressure Gauge.
Please make the CFL 1/2/3 before test. Do not skip this step.
1. Check the connection of RSP adapter
2. Check the cable of RSP (cut/ damaged cable)